GPS

DGC MEDICAL MAGNETIC RESONANCE IMAGING SYSTEM

OPER 0.4

OPER 0.35

OPER 0.3                                              OPER 0.23

    

Technical Data

Magnet Subsystem

Magnet Type: Open type permanent magnet (NdFeB) with vertical magnetic field and AMTC (adaptive magnet temperature control ) technique  

Magnetic Field: 0.5T, 0.4T, 0.35T, 0.3T, 0.23T

Opening of the Gantry: ≥ 300º    

Homogeneity: 20cm DSV <± 2ppm (FWHH) 

                         40cm DSV<5ppm(Vrms)

Patient Gap: 40cm  

Gradient Subsystem  

Amplifler: Global purchased high performance OEM product    

Gradient: 15~25 mT/m (Single Axis),26~43 mT/m (Effective)  

Slew Rate: 48~70 mT/m/ms (Single Axis),83~120 mT/m/ms (Effective)  

Nonlinearity: In a Ellipsoid of 40cm×40cm×38cm:<5%  

RF Subsystem  

Type of Transmitter and Receiver: Digital 

Receiver Channels: 4 channels,2 channels,1 channels  

Frequency of Transmitter and Receiver: DC~30MHz   

Max. Gain of Preamplifier: 30dB  

Factor of Noise: 0.3dB

Amplifier: 6KW (Global Purchased High Performance OEM Product) 

RF transmit coil: Flat  

Receive Coils: Phase array coils,quadrature coils ,linear coils 

Auto Prescanning Functions  

Auto RF correction

Auto Frequency Lock    

Auto Gain Adjustment 

Auto Recognition of Coils    

Auto Tuning of Coils   

Auto Linear Shimming Compensation  

Auto High Order Shimming Compensation

Auto Phase Correction  

Image Features  

Image Type: T1Weighted, T2 Weighted, T2* Weighted, Proton Density Weighted, Diffusion, MRA, Water Suppressed, Fat Suppressed  

Max.FOV: 400mm

Acquisition Matrix: 64×64~1024×1024 

Max. Display Martrix: 1024×1024

Spatial Resolution: 1mm (Head, Normal Model ) 

                                   1.5mm (Body, Normal Model)   

                                  0.5mm(Head, High Resolution Model)  

                                   0.75mm (Body, High Resolution Model)  

Slice Orientation: Transversal, Sagittal, Coronal, Oblique, Multi-slice & multi-angle 

Pulse Sequences

Spin Echo (SE) 

Multi Slice Multi Echo (MSME)   

Fast Spin Echo (FSE) 

Fast Revovery Fast Spin Echo (FRFSE)    

Single Shot Fast Spin Echo (SSFSE)

Multi Shot Fast Spin Echo (MSFSE)

Multi Shot Fast Spin Echo (MSFSE)    

Inversion Recovery (IR) 

Short Time Inversion Recovery (STIR) 

Fluid Attenuation Inversion Recovery (FLAIR)   

Inversion Recovery Fast Spin Echo (IRFSE)

2D Gradient Echo (2D GRE)    

3D Gradient Echo (3D GRE)   

Fast Gradient Echo 

Turtbo Gradient Echo  

Steady State Process Gradient Echo (SSPGRE)    

2D MRA (2D TOF) 

3D MRA (3D TOF ) / (MOTSA)  

Spin Echo Plannar Imaging (SEEPI)

Linear Scan Diffusion Imaging (LSDI)    

Fat-water Separation Imaging (DIXON)  

Rotate Acquisition Correction (RAC) 

Special Techniques 

Water Suppression

Fat Suppression    

Water Imaging (MRCP, MRU, MRM, etc)    

Variable Phase Gradient Encoding  

Magnetization Transfer Contrast (MTC)  

Flow Compensation    

Presaturation Tracing   

Over Sampling    

Breath-hold Imaging 

Partial Acquisition 

Parallet Acquisition    

Optimized Bandwidth Acquisition 

Maximum Intensity Projection (MIP)

Tilted Optimized Nonsaturation Excitation (TONE) 

ECG Gating    

Respiratory Gating   

Active Dynamic Shimming   

Motion Artifact Suppression   

Metal Artifact Suppression and Acquisition Optimization   

Elliptical K-space Coverage

Navigator Echo  

Shared Echo  

Patient Monitoring and Management    

Inner Communication  

Physiological Signal Monitoring System  

Patient Database

OPER 0.5

Patient Table  

3D Motorized Patient Table, Manual Operated Patient Table 

Load Capacity: 240Kg  

Image Display and Post-processing

Image Post-processing: Filter, Enhancement, Magnification, Cutting, Annotation, Image Translation, Auto Adjustment of Window Width and Window Centre  

Image Analysis: ROI, Distance Measurement, Area Calculation, Histogram, Distribution of Gray Scale  

Computer System  

Host: Professional Image Workstation with Dual Processors 

Clock Speed: ≥ 2.8 GHz 

Network Interface Speed: 1000M    

Memory: 2048MB   

Graphic Memory: 128MB    

Hard Disk: 160GB    

Compact Disk: CRW Writer or DVD Writer  

Image Reconstruction Speed: 285 images/s (256×256 matrix)  

Image Display: 20" TFT (Max. Resolution: 1280×1027)  

Laser Printer Interface  

Digital Interface  

DICOM 3.0 Compatible    

Image Transfer    

Standard DICOM 3.0 Format

Compatible with PACS 

© 2009 La Orden de los Dragones http://www.dragongroupchina.com/

  

  

CHINA POR CATALOGO